DRAFT FOR REVIEW — Not Doctor-approved for training. VERIFY items must be cleared before competency use.
Denture Try-In
SprintRay Try-In 2 — Lab Manufacturing Procedure
Smile Avenue Family Dentistry — In-House Laboratory
SprintRay Try-In 2 — Lab Manufacturing Procedure
|
|
| Document ID |
SOP-LAB-120 |
| Version |
0.1 DRAFT |
| Effective (target) |
09/10/2026 |
| Owner |
Lab Technician (process owner) |
| Status |
DRAFT FOR REVIEW |
Start: Case received in Medit Lab
End: Try-in bagged and logged in SEAZONA for outbound delivery
1. Purpose
This procedure standardizes in-house manufacturing of SprintRay Try-In 2 devices used to evaluate bite registration and occlusion for fully and partially edentulous patients before definitive denture fabrication.
2. Scope
In scope: Medit Lab intake → scan QC → design (VERIFY) → RayWare / Pro 2 → ProWash S → NanoCure → finish clean → bag & SEAZONA.
Out of scope: Definitive printed dentures (SOP-LAB-090); Apex Flex RPD (SOP-LAB-030); OnX Tough 2 (SOP-LAB-040); clinical bite evaluation appointment itself; non-SprintRay printers.
3. Roles
| Role |
Responsibility |
| Clinical Assistant / Office |
Sends case; receives bagged try-in |
| Lab Technician |
Scan QC; design approve; manufacture; bag; SEAZONA |
| Trainer / Lead Lab Technician |
Competency sign-off |
| Doctor |
Indication; clinical try-in evaluation; remake authorization |
4. Practice Configuration
| # |
Item |
Smile Avenue setting |
| 1 |
Printer |
Pro 2 only |
| 2 |
Cure unit |
NanoCure — Try-In 2 preprogrammed profile |
| 3 |
Case intake |
Medit Lab |
| 4 |
Design |
VERIFY — outside CAD or SprintRay Cloud Design |
| 5 |
Wash |
ProWash S — IPA ≥ 91% — standard preprogrammed cycle |
| 6 |
Layer thickness |
Typically 100 µm |
| 7 |
Orientation |
Intaglio facing away from platform |
| 8 |
Design approval before print |
Lab Technician |
| 9 |
Case log |
SEAZONA |
Thickness minima (IFU-006)
| Area |
Maxillary |
Mandibular |
| Lingual ridge |
≥ 2.0 mm |
≥ 2.0 mm |
| Palatal / lingual |
≥ 2.5 mm |
≥ 1.5 mm |
| Facial / buccal |
≥ 1.5 mm |
≥ 1.5 mm |
| Implant overdenture area |
≥ 2.0 mm |
≥ 2.0 mm |
5. Safety
- Treat uncured Try-In 2 as sensitizer. PPE + ventilation.
- Do not bag under-cured try-ins. Do not use expired resin.
6. Materials & Equipment
- SprintRay Try-In 2; IPA ≥ 91%
- Stack: Medit Lab → design (VERIFY) → RayWare → Pro 2 → ProWash S → NanoCure Try-In 2 → SEAZONA
7. Process Controls
| Gate |
Do not proceed if |
Proceed when |
| Design |
Below minima / not approved |
QC-A pass |
| Printer |
Not Pro 2 |
Pro 2 |
| Pre-cure |
Wet |
Fully dry |
| Cure |
Wrong profile |
Try-In 2 profile complete |
| Outbound |
Failed QC-E |
Bagged + SEAZONA |
8. Procedure
A. Intake & scan QC
- Medit Lab: verify ID, arch, edentulous status, due date.
- QC-SCAN.
B. Design (VERIFY)
- Design/receive try-in STL meeting thickness minima.
- Lab Technician approves. QC-A.
C–D. Nest & print
- Try-In appliance; Try-In 2 material; intaglio away; ~100 µm; Pro 2.
- Shake 1 minute; print; log lot. QC-B.
E–F. Wash & cure
- ProWash S standard cycle; dry. QC-C.
- NanoCure Try-In 2 profile; remove supports. QC-D.
G. Finish, clean, bag
- Soap + brush + warm water; dry; bag; SEAZONA. QC-E.
End of SOP-LAB-120. Clinical bite evaluation is Doctor-led outside this SOP.
9. Case Log (SEAZONA)
| Field |
Entry |
| Date / case ID / arch |
|
| Scan QC |
☐ Pass ☐ Fail |
| Design approved by (Lab Technician) |
|
| Try-In 2 lot |
|
| NanoCure Try-In 2 |
☐ Yes |
| Bagged outbound |
☐ Yes — time __ |
10. Quality Checkpoints
| ID |
When |
Pass |
Fail action |
| QC-SCAN |
After Medit Lab intake |
Scans usable |
Return to office |
| QC-A |
Pre-print |
Approved + thickness minima |
Return to design |
| QC-B |
Post-print |
Complete geometry |
Remake print |
| QC-C |
Post-wash |
Clean + dry |
Re-wash/dry |
| QC-D |
Post-cure |
Fully cured |
Remake / salvage |
| QC-E |
Before outbound |
Finish + clean + bag + SEAZONA |
Hold release |
11. Remake Criteria
| Finding |
Action |
| Failed scan QC |
Do not manufacture |
| Incomplete print / delamination |
Remake |
| Below thickness minima |
Remake from corrected design |
| Wet / under-cured / wrong Try-In 2 profile |
Hold; correct or remake |
| Light support nubs only |
Salvage — re-check QC-E |
| Unsure |
Escalate to Doctor / lab lead |
12. Competency
| Requirement |
Sign-off |
| Read SOP-LAB-120 |
Tech _ Date ___ |
| Demonstrates Medit Lab → bagged Try-In 2 outbound |
Trainer _ Date ___ |
| Completes 3 observed cases |
1. __ 2. _ 3. ___ |
| Trainer attests ready |
Trainer _ Date ___ |
| Lab lead authorizes unsupervised production |
_ Date ___ |
13. Troubleshooting & Escalation
| Problem |
Action |
| Bad scans |
Return to office |
| Soft after cure |
Confirm dry + Try-In 2 NanoCure profile |
| Misrouted as definitive denture |
Stop; use SOP-LAB-090 for definitive |
SprintRay Support: 1-800-914-8004.
14. References
- SprintRay Try-In 2 IFU-006_8 (effective 04/02/2025)
- Local IFU archive:
TryIn2.txt
- Related definitive path: SOP-LAB-090
Smile Avenue Family Dentistry — In-House Laboratory · Internal use only
SOP-LAB-120 DRAFT FOR REVIEW · Target effective 09/10/2026 · Supersedes prior draft revisions
Review annually or when SprintRay IFU / equipment changes. Confirm all VERIFY items before training use.