DRAFT FOR REVIEW — Not Doctor-approved for training. VERIFY items must be cleared before competency use.
Smile Avenue Family Dentistry
IN-HOUSE LABORATORY Cypress & Katy, TX
Document ID: SOP-LAB-090
Version 0.1 DRAFT  |  Target effective 09/10/2026

Printed Denture

SprintRay Apex Base / High Impact + teeth — Lab Manufacturing Procedure

Smile Avenue Family Dentistry — In-House Laboratory
SprintRay Apex Base / High Impact Denture Base + Apex Teeth / High Impact Teeth — Lab Manufacturing Procedure

Document ID SOP-LAB-090
Version 0.1 DRAFT
Effective (target) 09/10/2026
Owner Lab Technician (process owner)
Status DRAFT FOR REVIEW

Start: Case received in Medit Lab
End: Denture bagged and logged in SEAZONA for outbound delivery


1. Purpose

This procedure standardizes in-house manufacturing of SprintRay rigid printed removable dentures (base + teeth) using Apex Base or High Impact Denture Base with matching teeth resins, so every case meets the same quality, safety, and traceability standard before leaving the lab.

2. Scope

In scope: Medit Lab intake → scan QC → design (VERIFY path) → RayWare / Pro 2 (± Duo Kit) → ProWash S → assemble & NanoCure → finish & polish → lab clean → bag & SEAZONA.

Out of scope:
- Apex Flex RPD flexible partials → SOP-LAB-030 (different resin, NanoCure Apex Flex Denture, clasp workflow)
- OnX Tough 2 fixed hybrids / monolithic tooth-shade prosthetics → SOP-LAB-040 (not a separate pink base + teeth assemble path)
- Chairside try-in delivery; Try-In 2 interim → SOP-LAB-120; clinical disinfection; non-SprintRay printers

Contrast (use the correct SOP)

Path Resin concept Typical use NanoCure (known)
SOP-LAB-090 (this) Pink rigid base (Apex Base or High Impact Base) + printed teeth Removable full/partial denture baseplates Conventional Denture (MWG Removable Dentures)
SOP-LAB-030 Apex Flex flexible base + Apex Teeth Flexible RPD / clasps Apex Flex Denture
SOP-LAB-040 OnX Tough 2 ceramic-hybrid tooth shade Fixed hybrids / related OnX indications OnX Tough 2 profiles (e.g. Max Strength per hybrid MWG) — not this assemble path

3. Roles

Role Responsibility
Clinical Assistant / Office Sends case via Medit Lab; receives bagged denture
Lab Technician Scan QC; design approve; Duo/print; wash; assemble; cure; finish; bag; SEAZONA
Trainer / Lead Lab Technician Competency observation and sign-off
Doctor Indication, shade Rx, try-in, remake authorization

4. Practice Configuration

# Item Smile Avenue setting
1 Printer Pro 2 (Duo Kit preferred when available)
2 Base resin default VERIFY Apex Base vs High Impact Denture Base per Rx / stock
3 Teeth resin Apex Teeth (default A2 unless Rx) or High Impact Denture Teeth per Rx pairing
4 Cure unit NanoCure — Conventional Denture (MWG Removable Dentures for assembled Apex path); if on-screen name differs for High Impact pairing, VERIFY exact profile name — never invent minutes
5 Case intake Medit Lab
6 Design VERIFY — outside CAD or SprintRay Cloud Design per IFU
7 Wash ProWash S — IPA ≥ 91% — standard / multi-cycle per IFU
8 Bond Use denture base resin in sockets; roughen as needed; tack-cure before final NanoCure
9 Design approval before print Lab Technician
10 Case log SEAZONA

Apex Base thickness minima (IFU-021)

Area Max / Mand
Palatal / lingual ≥ 2.0 mm
Facial / buccal ≥ 2.0 mm
Implant overdenture area ≥ 2.5 mm

High Impact Base thickness minima (IFU-013)

Area Max / Mand
Lingual ridge ≥ 2.5 mm
Palatal / lingual ≥ 2.0 mm
Facial / buccal ≥ 2.0 mm
Implant overdenture area ≥ 2.5 mm

Teeth minimum thickness: ≥ 0.5 mm (Apex Teeth / HI Teeth IFUs).

5. Safety

6. Materials & Equipment

Resins (select per Rx)

Component Options
Base Apex Base (Light Pink, Original Pink, Dark reddish Pink, Opaque Original Pink) or High Impact Denture Base (Original Pink, Dark Pink, Original Meharry, Dark Meharry, Bubblegum Pink)
Teeth Apex Teeth (Bleach, A1, A2, B1) or High Impact Denture Teeth (Bleach, A1, A2, B1)
Bond Matching denture base resin into sockets

Equipment stack

Step System
Intake Medit Lab
Design VERIFY
Nest / print RayWare → Pro 2 + Duo Kit (preferred)
Wash ProWash S (+ Duo wash adapter if Duo)
Cure NanoCure — Conventional Denture (VERIFY if HI pairing differs)
Case log SEAZONA

Polish (IFU / MWG)

Scotch-Brite / Fuzzies → pumice + compound + muslin → optional pink compound + cotton buff; MWG also references Resilience polish sequence for removable dentures — follow practice polish kit.

7. Process Controls

Gate Do not proceed if Proceed when
Wrong product family Flex or OnX Tough 2 Rx on this SOP Route to SOP-LAB-030 or 040
Design Below thickness minima; not approved QC-A pass
Printer Base not on Pro 2 Pro 2
Pre-assemble Wet parts Fully dry
Cure Wrong profile / invented times Conventional Denture (or VERIFY HI) complete
Outbound Failed QC-E Bagged + SEAZONA

8. Procedure

A. Intake & scan QC

  1. Open Medit Lab; verify ID, arch, base resin family, shades, due date. Confirm not Flex RPD or OnX hybrid unless misrouted.
  2. Scan QC; QC-SCAN.

B. Design (VERIFY)

  1. Design/receive separate base + teeth STLs meeting thickness minima.
  2. Lab Technician approves. QC-A: minima met; shades correct; STLs labeled.

C. Nest (RayWare + Duo preferred)

  1. Duo Mode on Pro 2 when available:
    - Base: intaglio toward platform; ~60°; posterior closer; 100 µm.
    - Teeth: occlusal toward/parallel platform; intaglio away; 100 µm.
  2. Fallback: sequential base then teeth jobs — document in SEAZONA.

D. Print

  1. Shake each resin 1 minute; pour; assign tanks; print; log lots.
    QC-B: Complete geometry for base and teeth.

E. Wash & dry

  1. ProWash S with Duo adapter if used; IPA ≥ 91%; standard/multi-cycle per IFU.
  2. Dry fully — especially sockets.
    QC-C: Clean + dry.

F. Assemble & post-cure

  1. Roughen sockets/tooth bases if needed.
  2. Place drop of base resin in each socket; seat teeth; press firmly; tack-cure until stable.
  3. Place assembled denture (dry) in NanoCure; select Conventional Denture; follow on-screen prompts. If mixing resin lines, MWG directs Conventional Denture path. Do not invent times.
    QC-D: Bonded + fully cured.

G. Finish, clean, bag

  1. Remove supports; smooth; polish per kit; soap/steamer clean; dry.
  2. Bag + SEAZONA.
    QC-E: Required for release.

End of SOP-LAB-090. Clinical try-in/delivery outside this SOP (see also SOP-LAB-120 for Try-In 2).

9. Case Log (SEAZONA)

Field Entry
Date
Patient / case ID
Base resin + shade + lot ☐ Apex Base ☐ High Impact Base
Teeth resin + shade + lot ☐ Apex Teeth ☐ HI Teeth
Duo used ☐ Yes ☐ Sequential fallback
NanoCure profile Conventional Denture ☐ / other VERIFY: ____
Design approved by (Lab Technician)
QC-SCAN + A–E ☐ Yes
Bagged for outbound ☐ Yes — time __

10. Quality Checkpoints

ID When Pass Fail action
QC-SCAN Intake Usable Return
QC-A Pre-print Minima + approved Redesign
QC-B Post-print Complete Remake
QC-C Post-wash Clean + dry Re-wash/dry
QC-D Post-cure Bonded + cured Remake / re-bond
QC-E Outbound Finish + bag + SEAZONA Hold

11. Remake Criteria

Finding Action
Flex/OnX misroute Stop; correct SOP
Incomplete print / delamination Remake
Teeth debond Re-bond + re-cure if appropriate; else remake
Under-cured / wrong profile Hold; correct or remake
Below thickness minima printed Remake from corrected design
Unsure Escalate

12. Competency

Requirement Sign-off
Read SOP-LAB-090; distinguish vs 030 and 040 Tech _ Date ___
Demonstrates Duo assemble + Conventional Denture cure Trainer _ Date ___
3 observed cases to bagged outbound 1. __ 2. _ 3. ___
Unsupervised authorization _ Date ___

13. Troubleshooting & Escalation

Problem Action
Soft after cure Confirm dry + Conventional Denture profile
Debond Re-bond with base resin / remake
Wrong pink family used Doctor; remake if required

SprintRay Support: 1-800-914-8004.

14. References


Smile Avenue Family Dentistry — In-House Laboratory · Internal use only
SOP-LAB-090 DRAFT FOR REVIEW · Target effective 09/10/2026 · Supersedes prior draft revisions
Review annually or when SprintRay IFU / equipment changes. Confirm all VERIFY items before training use.