DRAFT FOR REVIEW — Not Doctor-approved for training. VERIFY items must be cleared before competency use.
Smile Avenue Family Dentistry
IN-HOUSE LABORATORY Cypress & Katy, TX
Document ID: SOP-LAB-080
Version 0.1 DRAFT  |  Target effective 09/10/2026

Direct-Print Retainers

SprintRay Retainer — Lab Manufacturing Procedure

Smile Avenue Family Dentistry — In-House Laboratory
SprintRay Retainer — Lab Manufacturing Procedure

Document ID SOP-LAB-080
Version 0.1 DRAFT
Effective (target) 09/10/2026
Owner Lab Technician (process owner)
Related Station Job Aid SOP-LAB-080-JA
Status DRAFT FOR REVIEW

Start: Case received in Medit Lab with scan QC
End: Retainer bagged and logged in SEAZONA for outbound delivery


1. Purpose

This procedure standardizes in-house manufacturing of SprintRay direct-print retainers so every case meets the same quality, safety, and traceability standard before leaving the lab.

2. Scope

In scope: Medit Lab intake → scan QC → design (SprintRay Cloud Design AI Retainer preferred per IFU; VERIFY practice path) → RayWare / Pro 2 → ProWash S → NanoCure → finish → lab clean → bag & SEAZONA.

Out of scope: Chairside scanning; clinical delivery/adjust; vac-form retainers on printed models (see SOP-LAB-110 for models); nightguards (SOP-LAB-020); non-SprintRay printers.

3. Roles

Role Responsibility
Clinical Assistant / Office Sends case via Medit Lab; receives bagged retainer
Lab Technician Scan QC; design import/export/approve; manufacture; finish; bag; SEAZONA
Trainer / Lead Lab Technician Competency observation and sign-off
Doctor Indication, Rx, try-in/delivery, remake authorization

4. Practice Configuration

# Item Smile Avenue setting
1 Printer Pro 2 only
2 Cure unit NanoCure — Retainer preprogrammed profile (do not invent minutes); allow cool before remove
3 Case intake Medit Lab
4 Design VERIFY — SprintRay Cloud Design AI Retainer preferred (IFU); outside CAD STL also acceptable
5 Wash ProWash S — IPA ≥ 91% — standard cleaning cycle; basket wash with intaglio facing propeller preferred
6 Design QC thickness ≥ 0.6 mm (IFU)
7 Lab clean before bag Soap + brush + warm water; dry. Cleaning — not chemical disinfection.
8 Design approval before print Lab Technician
9 Case log SEAZONA

5. Safety

6. Materials & Equipment

Resins

Equipment stack

Step System
Intake Medit Lab
Design VERIFY — Cloud Design AI Retainer or outside CAD
Nest / print RayWare → Pro 2
Wash ProWash S
Cure NanoCure — Retainer
Case log SEAZONA

Design QC targets

Area Minimum
Appliance thickness ≥ 0.6 mm

7. Process Controls

Gate Do not proceed if Proceed when
Intake Missing / unreadable Case open with ID / Rx
Scan QC Holes / wrong arch Scans usable
Printer Not Pro 2 Pro 2 only
Design Not approved; < 0.6 mm QC-A pass
Resin Expired In-date; shaken 1 min
Pre-cure Wet Fully dry
Cure Wrong profile / hot remove deforming Retainer profile complete + cooled
Outbound Failed QC-E Bagged + SEAZONA

8. Procedure

A. Intake & scan QC

  1. Open case in Medit Lab.
  2. Verify ID, arch, Rx notes, due date.
  3. Scan QC; QC-SCAN pass/fail.

B. Design (VERIFY path)

  1. Preferred per IFU: SprintRay Cloud Design → AI Retainer treatment; upload scans; receive design. Alternately import outside CAD STL.
  2. Export STL; Lab Technician approves.
    QC-A: ≥ 0.6 mm; coverage matches Rx; STL labeled.

C. Nest (RayWare)

  1. Import STL; select Retainer appliance type; material Retainer; Pro 2.
  2. Orientation: intaglio away; ~10° from anterior; layer thickness typically ~170 µm default; default supports.
  3. Queue to Pro 2.

D. Print

  1. Shake 1 minute; pour; assign material; print; log lot.
    QC-B: Fully formed; no delamination.

E. Wash & dry

  1. ProWash S standard cycle, IPA ≥ 91%; prefer basket with intaglio toward propeller; remove promptly; dry fully (compressed air).
    QC-C: Clean + dry.

F. Post-cure

  1. NanoCure Retainer profile. Allow to cool before removing from device.
  2. Prefer support removal after cool cure; silicone bur to smooth.
    QC-D: Cured; shape intact; no tack.

G. Finish, clean, bag

  1. Optional: thin mineral oil on occlusal / support nubs + cotton buff for sheen/clarity.
  2. Soap + brush + warm water; dry.
  3. Bag + label + SEAZONA outbound.
    QC-E: Pass required for release.

End of SOP-LAB-080. Clinical delivery is outside this procedure.

9. Case Log (SEAZONA)

Field Entry
Date
Patient / case ID
Arch ☐ Max ☐ Mand ☐ Both
Scan QC ☐ Pass ☐ Fail
Design path ☐ Cloud Design AI ☐ Outside CAD (VERIFY)
Design approved by (Lab Technician)
Retainer lot
Printer Pro 2 ☐
NanoCure profile Retainer ☐
QC-SCAN + A–E ☐ Yes
Bagged for outbound ☐ Yes — time __

10. Quality Checkpoints

ID When Pass Fail action
QC-SCAN Intake Usable scans Return
QC-A Pre-print ≥ 0.6 mm + approved Redesign
QC-B Post-print Complete Remake
QC-C Post-wash Clean + dry Re-wash/dry
QC-D Post-cure Cured + cooled shape OK Remake / salvage
QC-E Outbound Finish + bag + SEAZONA Hold

11. Remake Criteria

Finding Action
Failed scan QC Do not manufacture
Incomplete / delamination Remake
Deformed from hot handling Remake
Under-cured / wrong profile Hold; remake if needed
Light nubs only Salvage
Unsure Escalate

12. Competency

Requirement Sign-off
Read SOP-LAB-080 and JA Tech _ Date ___
Demonstrates intake through bagged outbound Trainer _ Date ___
3 observed cases 1. __ 2. _ 3. ___
Unsupervised authorization _ Date ___

13. Troubleshooting & Escalation

Problem Action
Cloudy after deburr Mineral oil coat (optional per IFU)
Soft / tacky Confirm dry + Retainer NanoCure profile
Deformed Likely removed hot — remake; cool fully next time

SprintRay Support: 1-800-914-8004.

14. References


Smile Avenue Family Dentistry — In-House Laboratory · Internal use only
SOP-LAB-080 DRAFT FOR REVIEW · Target effective 09/10/2026 · Supersedes prior draft revisions
Review annually or when SprintRay IFU / equipment changes. Confirm all VERIFY items before training use.