DRAFT FOR REVIEW — Not Doctor-approved for training. VERIFY items must be cleared before competency use.
Smile Avenue Family Dentistry
IN-HOUSE LABORATORY Cypress & Katy, TX
Document ID: SOP-LAB-070
Version 0.1 DRAFT  |  Target effective 09/10/2026

Surgical Guides

SprintRay Surgical Guide 3 / Precision Guide — Lab Manufacturing Procedure

Smile Avenue Family Dentistry — In-House Laboratory
SprintRay Surgical Guide 3 (primary) / Precision Guide (Pro 2-only fork) — Lab Manufacturing Procedure

Document ID SOP-LAB-070
Version 0.1 DRAFT
Effective (target) 09/10/2026
Owner Lab Technician (process owner)
Status DRAFT FOR REVIEW

Start: Case received in Medit Lab with scan QC
End: Guide bagged and logged in SEAZONA for outbound delivery


1. Purpose

This procedure standardizes in-house manufacturing of SprintRay surgical guides — Surgical Guide 3 as the primary resin path, with a documented Precision Guide fork for Pro 2-only cases — so every guide meets the same quality, safety, and traceability standard before leaving the lab.

2. Scope

In scope: Medit Lab intake → scan QC → design (VERIFY path) → RayWare / Pro 2 print → ProWash S → NanoCure → finish → lab clean → bag & SEAZONA. Sleeve insertion and autoclave sterilization may be completed in-lab per IFU when practice policy requires (VERIFY who autoclaves — lab vs clinical).

Out of scope: Chairside scanning / CBCT planning technique; surgical placement; patient delivery; non-SprintRay printers.

3. Roles

Role Responsibility
Clinical Assistant / Office Sends case via Medit Lab; receives bagged guide
Lab Technician Scan QC; design approve; manufacture; finish; bag; SEAZONA; sleeve/autoclave if assigned
Trainer / Lead Lab Technician Competency observation and sign-off
Doctor Indication, implant plan approval, surgery, remake authorization

4. Practice Configuration

# Item Smile Avenue setting
1 Printer Pro 2 only
2 Primary resin Surgical Guide 3
3 Fork resin Precision Guide — Pro 2 only (IFU-validated solely on Pro 2 + NanoCure)
4 Cure unit NanoCure — Surgical Guide 3 or Precision Guide exact profile (do not invent minutes)
5 Case intake Medit Lab
6 Design VERIFY — outside CAD or SprintRay Cloud Design per IFU
7 Wash ProWash S — IPA ≥ 91% — Surgical Guide 3: standard preprogrammed cycle; Precision Guide: Standard Multi-Cycle 3/3/3
8 Sterilization (IFU) Insert sleeves → autoclave 5 min at 134 °C; Precision Guide also: dry 15 min at 134 °C. VERIFY whether lab or clinical performs autoclave.
9 Design approval before print Lab Technician
10 Case log SEAZONA

5. Safety

6. Materials & Equipment

Resins

Resin When
SprintRay Surgical Guide 3 Primary Smile Avenue path
SprintRay Precision Guide Fork when Rx specifies Precision Guide (Pro 2-only)

Also: IPA ≥ 91%; surgical guide sleeves per implant system (VERIFY sleeve inventory / Rx).

Equipment stack

Step System
Intake Medit Lab
Design VERIFY — outside CAD or SprintRay Cloud Design
Nest / print RayWare / Dashboard Print Setup → Pro 2
Wash ProWash S
Cure NanoCure — exact resin profile
Sterilize Autoclave per IFU (VERIFY ownership)
Case log SEAZONA

Design QC

Confirm sleeve holes, seating, and plan match approved surgical Rx; STL labeled; correct resin family selected.

7. Process Controls

Gate Do not proceed if Proceed when
Intake Case missing / unreadable Case open with ID / Rx / plan
Scan QC Inadequate anatomy / wrong files Scans usable
Resin fork Precision Guide queued off Pro 2 Surgical Guide 3 or Precision Guide on Pro 2
Design Not Lab Technician–approved QC-A pass
Resin Expired / wrong family In-date correct resin; shaken 1 min
Pre-cure Part wet Fully dry
Cure Wrong profile / invented times Exact NanoCure profile complete
Outbound Failed QC-E; not bagged/logged Bagged + SEAZONA ready

8. Procedure

A. Intake & scan QC

  1. Open case in Medit Lab.
  2. Verify ID, arch, implant sites, resin (Surgical Guide 3 vs Precision Guide), sleeve system, due date.
  3. Scan QC: required scans/models usable; no critical holes or wrong-patient files.
  4. QC-SCAN: Pass → continue. Fail → return to office.

B. Design (VERIFY path)

  1. Import into practice design path (VERIFY: outside CAD or SprintRay Cloud Design).
  2. Design/receive surgical guide STL per approved plan.
  3. Export STL; Lab Technician approves.
    QC-A: Holes/sleeve positions match plan; seating OK; correct resin noted; STL labeled. Fail → redesign.

C. Nest (RayWare / Print Setup)

  1. Import approved STL.
  2. Select appliance type Surgical Guide; printer Pro 2.
  3. Select Surgical Guide 3 or Precision Guide per Rx.
  4. Orientation: intaglio facing away; layer thickness typically 100 µm; default / auto supports (Precision Guide: Dashboard Print Setup auto-orient).
  5. Queue to Pro 2 only.

D. Print

  1. Shake resin 1 minute; pour to min fill; record lot; assign material; clean platform; start print.
  2. QC-B: Fully formed; sleeve holes open/usable; no delamination. Fail → remake.

E. Wash & dry

  1. Surgical Guide 3: ProWash S standard preprogrammed cycle, IPA ≥ 91%.
    Precision Guide: ProWash S Standard Multi-Cycle 3/3/3.
  2. Fully dry.
    QC-C: Clean + dry.

F. Post-cure

  1. NanoCure: select Surgical Guide 3 or Precision Guide exact profile. Do not invent times.
  2. Remove supports (flush cutter / diamond disc); smooth with Scotch-Brite / Fuzzies. Optional pumice/compound polish.
    QC-D: Fully cured; geometry intact.

G. Finish, sleeves, sterilize (VERIFY ownership), bag

  1. Lab clean with soap + brush + warm water; dry.
  2. Insert surgical guide sleeves into drill holes when assigned to lab.
  3. If lab autoclaves: 5 min at 134 °C (Surgical Guide 3); Precision Guide: 5 min at 134 °C then dry 15 min at 134 °C. Note expected color change after autoclave.
  4. If clinical autoclaves: bag as clean finished guide with sleeves status noted in SEAZONA; do not claim sterile unless autoclaved.
  5. Label, bag, SEAZONA bagged / outbound.
    QC-E: Finish OK; cleaned; bagged; logged; sterilization status documented.

End of SOP-LAB-070. Surgery and clinical delivery are outside this procedure.

Precision Guide fork summary (Pro 2-only)

Item Precision Guide
Printer Pro 2 only (IFU)
Wash 3/3/3 multi-cycle
Cure NanoCure only (IFU for Pro 2)
Autoclave 5 min @ 134 °C + 15 min dry @ 134 °C

9. Case Log (SEAZONA)

Field Entry
Date
Patient / case ID
Medit Lab reference
Resin ☐ Surgical Guide 3 ☐ Precision Guide
Scan QC ☐ Pass ☐ Fail
Design approved by (Lab Technician)
Resin lot
Printer Pro 2 ☐
NanoCure profile Surgical Guide 3 ☐ / Precision Guide ☐
Sleeves inserted ☐ Yes ☐ No — who: ____
Autoclave ☐ Lab ☐ Clinical ☐ N/A — cycle logged: ____
QC-SCAN + A–E ☐ Yes
Bagged for outbound ☐ Yes — time __

10. Quality Checkpoints

ID When Pass Fail action
QC-SCAN Intake Scans usable Return to office
QC-A Pre-print Plan match + approved Redesign
QC-B Post-print Complete + holes OK Remake
QC-C Post-wash Clean + dry Re-wash/dry
QC-D Post-cure Cured + intact Remake / salvage
QC-E Outbound Finish + bag + SEAZONA + sterilize status Hold

11. Remake Criteria

Finding Action
Failed scan QC Do not manufacture
Incomplete print / closed sleeve holes / delamination Remake
Precision Guide off Pro 2 Stop; remake on Pro 2
Under-cured / wrong profile Hold; correct or remake
Crack after autoclave Remake
Unsure Escalate to Doctor / lab lead

12. Competency

Requirement Sign-off
Read SOP-LAB-070 Tech _ Date ___
Demonstrates both Surgical Guide 3 and Precision Guide fork controls Trainer _ Date ___
Completes 3 observed cases to bagged outbound 1. __ 2. _ 3. ___
Trainer attests ready Trainer _ Date ___
Lab lead authorizes unsupervised production _ Date ___

13. Troubleshooting & Escalation

Problem Action
Bad scans / missing plan Return to office / Doctor
Yellow after cure (SG3) Expected; may clear after sterilization
Soft after cure Confirm dry + correct NanoCure profile
Precision Guide requested on non–Pro 2 Stop; use Pro 2 or switch to Surgical Guide 3 per Doctor

SprintRay Support: 1-800-914-8004.

14. References


Smile Avenue Family Dentistry — In-House Laboratory · Internal use only
SOP-LAB-070 DRAFT FOR REVIEW · Target effective 09/10/2026 · Supersedes prior draft revisions
Review annually or when SprintRay IFU / equipment changes. Confirm all VERIFY items before training use.