SprintRay Ceramic Crown — Lab Manufacturing Procedure
Smile Avenue Family Dentistry — In-House Laboratory
SprintRay Ceramic Crown — Lab Manufacturing Procedure
| Document ID | SOP-LAB-060 |
| Version | 0.1 DRAFT |
| Effective (target) | 09/10/2026 |
| Owner | Lab Technician (process owner) |
| Related | Station Job Aid SOP-LAB-060-JA |
| Status | DRAFT FOR REVIEW |
Start: Case received in Medit Lab with scan QC
End: Restoration bagged and logged in SEAZONA for outbound delivery
This procedure standardizes in-house manufacturing of SprintRay Ceramic Crown definitive restorations (full/partial crowns, veneers, and related IFU indications) so every case meets the same quality, safety, and traceability standard before leaving the lab.
In scope: Medit Lab intake → scan QC → design (VERIFY path) → RayWare / Pro 2 print → wash/dry → NanoCure → finish & polish → lab clean → bag & SEAZONA outbound log.
Out of scope: Chairside scanning; clinical try-in, cementation, and patient delivery; chairside chemical disinfection; temporary Digital Temp (SOP-LAB-050); Crown HT / Midas Restore / Motto paths; non-SprintRay printers.
IFU indications include: individual/fixed definitive full single crowns; definitive partial crowns (anterior/posterior); individual/fixed single veneers; artificial teeth for removable definitive full dentures; individual removable monolithic full and partial dentures (when prescribed as Ceramic Crown path).
| Role | Responsibility |
|---|---|
| Clinical Assistant / Office | Sends case via Medit Lab; receives bagged restoration from lab |
| Lab Technician | Scan QC; design import/export/approve; manufacture; finish; bag; SEAZONA log |
| Trainer / Lead Lab Technician | Competency observation and sign-off |
| Doctor | Clinical indication, shade/Rx, try-in/cementation, remake authorization (outside manufacturing steps) |
| # | Item | Smile Avenue setting |
|---|---|---|
| 1 | Printer | Pro 2 only for this path |
| 2 | Cure unit | NanoCure — select exact Ceramic Crown resin profile (on-screen name; do not invent minutes) |
| 3 | Case intake | Medit Lab |
| 4 | Design | VERIFY — outside CAD or SprintRay Cloud Design per IFU (Lab Technician imports, exports STL, approves) |
| 5 | Wash | ProWash S available — IPA ≥ 91%. IFU primary methods: IPA bowl scrub or spray/wipe until resin-free; dry completely before cure. VERIFY whether practice standardizes ProWash S cycle vs hand-wash for filled Ceramic Crown. |
| 6 | Tank preference | Optical Polish Tank preferred (IFU); Standard Tank acceptable |
| 7 | Lab clean before bag | Soap, toothbrush, and warm water (and/or steamer), then dry. Cleaning — not chemical disinfection. |
| 8 | Design approval before print | Lab Technician |
| 9 | Case log | SEAZONA |
| 10 | Design QC thickness | ≥ 0.6 mm (IFU) |
Contraindications: known allergy; intraoral contact with uncured resin; use outside IFU indications.
| Step | System |
|---|---|
| Intake | Medit Lab |
| Design | VERIFY — outside CAD or SprintRay Cloud Design |
| Nest / print | RayWare → Pro 2 |
| Wash | ProWash S and/or IFU IPA methods (VERIFY practice standard) |
| Cure | NanoCure — Ceramic Crown profile |
| Case log | SEAZONA |
| Area | Minimum |
|---|---|
| Restoration thickness | ≥ 0.6 mm |
Confirm shade, margins, contacts, and occlusion match Rx.
| Step | Tools / media |
|---|---|
| Support removal | Flush cutter or diamond disc |
| Polish | SprintRay Restorative Finishing Kit (preferred) |
| Optional characterization | OptiGlaze (GC) colors/stains + clear glaze |
| Post-cure wash | IPA bowl scrub or ProWash/Dry after cure (IFU) |
| Cleanup | Steamer and/or dish soap + brush + warm water; dry |
| Gate | Do not proceed if | Proceed when |
|---|---|---|
| Intake | Case missing or files unreadable | Case open with matching ID / Rx |
| Scan QC | Holes, missing prep/bite, inadequate margins | Scans usable for design |
| Printer | Not Pro 2 | Pro 2 only |
| Design | Not Lab Technician–approved; thickness < 0.6 mm | QC-A pass |
| Resin | Expired or wrong material | In-date Ceramic Crown; shaken 1 minute |
| Pre-cure | Part wet / residual liquid resin | Fully dry and resin-free |
| Cure | Wrong profile / custom times invented | NanoCure Ceramic Crown profile complete |
| Outbound | Failed QC-E; not bagged; not logged | Bagged + SEAZONA outbound ready |
End of SOP-LAB-060. Clinical cementation and delivery are outside this procedure.
| Field | Entry |
|---|---|
| Date | |
| Patient / case ID | |
| Medit Lab reference | |
| Teeth / type | ☐ Crown ☐ Veneer ☐ Other IFU: ____ |
| Scan QC | ☐ Pass ☐ Fail → returned |
| Designed by (VERIFY path) | |
| Design approved by (Lab Technician) | |
| Ceramic Crown shade + lot | |
| Printer | Pro 2 ☐ |
| Wash method | ☐ ProWash S ☐ IPA hand ☐ Both |
| NanoCure profile | Ceramic Crown ☐ |
| Lab Technician initials | |
| QC-SCAN + A–E pass | ☐ Yes |
| Remake? | ☐ No ☐ Yes — reason: __ |
| Bagged for outbound | ☐ Yes — time __ |
| ID | When | Pass | Fail action |
|---|---|---|---|
| QC-SCAN | After Medit Lab intake | Scans usable | Return to office |
| QC-A | Pre-print | Approved + ≥ 0.6 mm | Return to design |
| QC-B | Post-print | Complete geometry | Remake |
| QC-C | Post-wash | Clean + dry | Re-wash/dry |
| QC-D | Post-cure | Fully cured | Remake / salvage |
| QC-E | Before outbound | Finish + bag + SEAZONA | Hold |
| Finding | Action |
|---|---|
| Failed scan QC | Do not manufacture |
| Incomplete print / delamination | Remake |
| Printed on non–Pro 2 | Remake on Pro 2 |
| Wet / under-cured / wrong profile | Hold; correct or remake |
| Margin chip / critical fracture | Remake |
| Light support nubs only | Salvage → re-check QC-E |
| Unsure | Escalate to Doctor / lab lead |
| Requirement | Sign-off |
|---|---|
| Read SOP-LAB-060 and Job Aid SOP-LAB-060-JA | Tech _ Date ___ |
| Demonstrates Medit Lab intake, scan QC, design approve, process controls | Trainer _ Date ___ |
| Completes 3 observed cases to bagged outbound | 1. __ 2. _ 3. ___ |
| Trainer attests ready | Trainer _ Date ___ |
| Lab lead authorizes unsupervised production | _ Date ___ |
| Problem | Action |
|---|---|
| Bad scans | QC-SCAN fail → office |
| Print failure | Remake; filter; verify material assignment |
| Soft / under-cured | Confirm dry + Ceramic Crown NanoCure profile |
| Shade variance | Confirm shake/stir/cure; remake if needed |
| Allergy / wrong indication | Stop — Doctor |
SprintRay Support: 1-800-914-8004 — missing NanoCure Ceramic Crown profile; repeated failures.
/workspace/sprintray-ifu/all/CeramicCrown.txtSmile Avenue Family Dentistry — In-House Laboratory · Internal use only
SOP-LAB-060 DRAFT FOR REVIEW · Target effective 09/10/2026 · Supersedes prior draft revisions
Review annually or when SprintRay IFU / equipment changes. Confirm all VERIFY items before training use.