SprintRay Digital Temp — Lab Manufacturing Procedure
Smile Avenue Family Dentistry — In-House Laboratory
SprintRay Digital Temp — Lab Manufacturing Procedure
| Document ID | SOP-LAB-050 |
| Version | 0.1 DRAFT |
| Effective (target) | 09/10/2026 |
| Owner | Lab Technician (process owner) |
| Related | Station Job Aid SOP-LAB-050-JA |
| Status | DRAFT FOR REVIEW |
Start: Case received in Medit Lab with scan QC
End: Temporary restoration bagged and logged in SEAZONA for outbound delivery
This procedure standardizes in-house manufacturing of SprintRay Digital Temp temporary full/partial crowns and temporary bridges so every case meets the same quality, safety, and traceability standard before leaving the lab.
In scope: Medit Lab intake → scan QC → design (VERIFY path) → RayWare / Pro 2 print → ProWash S → NanoCure → finish & polish → lab clean → bag & SEAZONA outbound log.
Out of scope: Chairside scanning; clinical try-in, cementation, and patient delivery; chairside chemical disinfection; definitive Ceramic Crown (SOP-LAB-060); Crown HT / Midas Restore paths; non-SprintRay printers.
| Role | Responsibility |
|---|---|
| Clinical Assistant / Office | Sends case via Medit Lab; receives bagged temporary from lab |
| Lab Technician | Scan QC; design import/export/approve; manufacture; finish; bag; SEAZONA log |
| Trainer / Lead Lab Technician | Competency observation and sign-off |
| Doctor | Clinical indication, shade/Rx, try-in/cementation, remake authorization (outside manufacturing steps) |
| # | Item | Smile Avenue setting |
|---|---|---|
| 1 | Printer | Pro 2 only for this path |
| 2 | Cure unit | NanoCure — select exact Digital Temp resin profile (on-screen name; do not invent minutes) |
| 3 | Case intake | Medit Lab |
| 4 | Design | VERIFY — outside CAD or SprintRay Cloud Design per IFU (Lab Technician imports, exports STL, approves) |
| 5 | Wash | ProWash S — IPA ≥ 91% — Standard Multi-Cycle Wash (3 min wash / 3 min rinse / 3 min dry) preferred; or standard preprogrammed cycle |
| 6 | Shades | Bleach, A1, A2, B1 — select per prescription |
| 7 | Duration of use (IFU) | Expected ≤ 29 days |
| 8 | Lab clean before bag | Soap, toothbrush, and warm water (and/or steamer), then dry. Cleaning — not chemical disinfection. |
| 9 | Design approval before print | Lab Technician |
| 10 | Case log | SEAZONA |
| 11 | Design QC thickness | ≥ 0.6 mm (IFU) |
Contraindications: known allergy; intraoral contact with uncured resin; use outside temporary crown/bridge indications.
| Step | System |
|---|---|
| Intake | Medit Lab |
| Design | VERIFY — outside CAD or SprintRay Cloud Design |
| Nest / print | RayWare → Pro 2 |
| Wash | ProWash S |
| Cure | NanoCure — Digital Temp profile |
| Case log | SEAZONA |
| Area | Minimum |
|---|---|
| Restoration thickness | ≥ 0.6 mm |
Confirm shade, unit count, and margins match Rx; intaglio free of critical voids.
| Step | Tools / media |
|---|---|
| Smooth supports | Scotch-Brite / Fuzzies wheel |
| Polish | Pumice + polishing compound + muslin |
| Optional mirror | Pink compound + cotton buff |
| Optional characterization | IPS Empress Direct (Ivoclar) per workflow guide |
| Cleanup | Steamer and/or soap + brush + warm water; dry |
| Gate | Do not proceed if | Proceed when |
|---|---|---|
| Intake | Case missing or files unreadable in Medit Lab | Case open with matching ID / Rx |
| Scan QC | Holes, missing prep/bite, inadequate margins | Scans usable for design |
| Printer | Queued to any printer other than Pro 2 | Pro 2 only |
| Design | Not Lab Technician–approved; thickness < 0.6 mm | QC-A pass |
| Resin | Expired or wrong shade/family | In-date Digital Temp per Rx; shaken 1 minute |
| Wash | IPA < 91% | ProWash S cycle complete |
| Pre-cure | Part wet | Fully dry |
| Cure | Wrong profile / custom times invented | NanoCure Digital Temp profile complete |
| Outbound | Failed QC-E; not bagged; not logged | Bagged + SEAZONA outbound ready |
End of SOP-LAB-050. Office receipt, chairside disinfection, try-in, and cementation (e.g., Telio Link per IFU) are clinical — not this manufacturing procedure.
| Field | Entry |
|---|---|
| Date | |
| Patient / case ID | |
| Medit Lab reference | |
| Teeth / units | |
| Scan QC | ☐ Pass ☐ Fail → returned |
| Designed by (VERIFY path) | |
| Design approved by (Lab Technician) | |
| Digital Temp shade + lot | |
| Printer | Pro 2 ☐ |
| ProWash S cycle OK | ☐ Yes |
| NanoCure profile | Digital Temp ☐ |
| Lab Technician initials | |
| QC-SCAN + A–E pass | ☐ Yes |
| Remake? | ☐ No ☐ Yes — reason: __ |
| Bagged for outbound | ☐ Yes — time __ |
No SEAZONA bagged status = not released from the lab.
| ID | When | Pass | Fail action |
|---|---|---|---|
| QC-SCAN | After Medit Lab intake | Scans usable | Return to office |
| QC-A | Pre-print | Approved design + ≥ 0.6 mm | Return to design |
| QC-B | Post-print | Complete geometry | Remake print |
| QC-C | Post-wash | Clean + dry | Re-wash/dry |
| QC-D | Post-cure | Fully cured; supports clean | Remake / salvage |
| QC-E | Before outbound | Finish + clean + bag + SEAZONA | Hold release |
| Finding | Action |
|---|---|
| Failed scan QC | Do not manufacture — request new scans |
| Incomplete print / delamination / major voids | Remake print |
| Printed on non–Pro 2 | Remake on Pro 2 |
| Wet / under-cured / wrong cure profile | Hold outbound; correct or remake |
| Torn margins after supports | Remake |
| Light support nubs only | Salvage — finish and re-check QC-E |
| Unsure | Stop and escalate to Doctor / lab lead |
A Lab Technician may run this procedure unsupervised after:
| Requirement | Sign-off |
|---|---|
| Read SOP-LAB-050 and Job Aid SOP-LAB-050-JA | Tech _ Date ___ |
| Demonstrates Medit Lab intake, scan QC, design import/export/approve, and process controls | Trainer _ Date ___ |
| Completes 3 observed cases to bagged outbound with SEAZONA log | 1. __ 2. _ 3. ___ |
| Trainer attests ready | Trainer _ Date ___ |
| Lab lead authorizes unsupervised production | _ Date ___ |
| Problem | Action |
|---|---|
| Bad scans in Medit Lab | QC-SCAN fail → return to office |
| Print failure | Remake; filter resin; verify material assignment on Pro 2 |
| Soft / tacky after cure | Confirm dry part and Digital Temp NanoCure profile; remake if needed |
| Shade mismatch | Confirm shake/mix and cure; remake if Rx shade wrong |
| Allergy or wrong indication noted | Stop — Doctor; do not manufacture |
Escalate to Doctor / lab lead: unclear Rx; allergy; chronic remakes; IFU deviation requests.
Escalate to office (Medit Lab / SEAZONA): failed scan QC; missing prescription.
SprintRay Support: 1-800-914-8004 — missing NanoCure Digital Temp profile; repeated Pro 2 failures.