← Lab SOP Library · Smile Avenue Family Dentistry — Internal
Smile Avenue Family Dentistry
IN-HOUSE LABORATORY Cypress & Katy, TX
Document ID: SOP-LAB-030
Version 1.0  |  Effective 09/10/2026

Removable Partial Denture Fabrication

SprintRay Apex Flex & Apex Teeth — Lab Manufacturing Procedure

OwnerLab Technician (process owner)
Applies toSmile Avenue Family Dentistry in-house lab
StartCase received in Medit Lab
EndProsthetic bagged and logged in SEAZONA for outbound delivery
RelatedStation Job Aid SOP-LAB-030-JA

1. Purpose

This procedure standardizes in-house manufacturing of SprintRay flexible removable partial dentures (Apex Flex Partial Base with Apex Teeth) so every case meets the same quality, safety, and traceability standard before leaving the lab.

2. Scope

3. Roles

RoleResponsibility
Clinical Assistant / OfficeSends case via Medit Lab; receives bagged prosthetic from lab
Lab TechnicianScan QC, Heygears portal submit/export/approve, manufacture, finish, bag, SEAZONA log
Trainer / Lead Lab TechnicianCompetency observation and sign-off
DoctorClinical indication, try-in, and remake authorization (outside manufacturing steps)

4. Practice Configuration

#ItemSmile Avenue setting
1Printer (Flex base)Pro 2
2Cure unitNanoCure — cycle Apex Flex Denture
3Case intakeMedit Lab
4DesignHeygears portal (Heygears designs; Lab Technician imports, exports, approves)
5Duo KitYes
6WashProWash S
7Apex Flex shadesStock Standard Pink and Light Pink; select per prescription
8Apex Teeth defaultA2 (other shades per prescription)
9Lab clean before bagSoap, toothbrush, and warm water (and/or steamer), then dry. This is cleaning — not chemical disinfection.
10Design approval before printLab Technician
11Case logSEAZONA

5. Safety

6. Materials & Equipment

Resins

Equipment stack

StepSystem
IntakeMedit Lab
DesignHeygears portal
Nest / printRayWare → Pro 2 + Duo Kit
WashProWash S
CureNanoCure — Apex Flex Denture
Case logSEAZONA

Design QC targets (review Heygears design)

AreaMinimum
Palatal / lingual wall≥ 1.0 mm
Facial / buccal wall≥ 1.0 mm
Implant overdenture area≥ 1.5 mm
Apex Teeth thickness≥ 0.5 mm
Clasp thickness (target)≥ 1.2 mm

Export separate base and teeth STL files for Duo Kit printing.

Polish kit (SprintRay Master Workflow Guide — Partial Dentures)

StepTools / media
Smooth supportsLab handpiece; abrasive wheel or carbide bur (low RPM); Scotch-Brite / Fuzzies as needed
Resilience polishDental lathe; black bristle wheel B27/B29; Resilience polish (wet, not runny)
InterproximalsLab handpiece; Robinson #11; Tripoli Rouge
High shineDental lathe; cotton buff; Ivoclar Vivadent Universal Polishing Paste
Condition / cleanupMineral oil; fresh Robinson #11; steamer and/or soap + toothbrush + warm water
Technique: Polish at low speed. Do not let wheels dry out — heat can damage flexible resin.

7. Process Controls

GateDo not proceed ifProceed when
IntakeCase missing or files unreadable in Medit LabCase open with matching ID / Rx
Scan QCHoles, missing bite, inadequate soft tissueScans usable for design
PrinterFlex base queued to Pro / Pro SFlex base on Pro 2 only
DesignNot Lab Technician–approved; below minimaQC-A pass
ResinExpired or wrong materialIn-date Apex Flex + Apex Teeth; shaken 1 minute
WashIPA < 91%ProWash S standard multi-cycle
Pre-curePart wet; base supports already removedFully dry; base supports still on
CureWrong cycle / classic ProCure for Flex baseNanoCure Apex Flex Denture complete
OutboundFailed QC-E; not bagged; not loggedBagged + SEAZONA outbound ready

8. Procedure

A. Intake &amp; scan QC

1
Open the case in Medit Lab.
2
Verify patient/case ID, arch, and prescription notes (Flex shade, teeth shade, due date).
3
Perform scan QC: required arches captured; bite usable; soft tissue/ridge adequate; no critical holes or wrong-patient files.
4
QC-SCAN: Pass → continue. Fail → stop manufacturing; request rescan or clarified Rx from the office. Do not design on failing scans.

B. Design (Heygears portal)

5
Import accepted scans into the Heygears portal.
6
Heygears designs the flexible partial per prescription (clasps, teeth, shade targets).
7
Export print-ready STL files (base + teeth for Duo).
8
Lab Technician approves the design for print.
QC-A: Clasps ≥ 1.2 mm; walls meet minima; occlusion reasonable; shades correct; STLs labeled.
If QC-A fails — return to Heygears for redesign; do not nest.

C. Nest (RayWare + Duo Kit)

9
Import approved STL(s) into RayWare.
10
Select Duo Mode on Pro 2.
11
Tank I: Partial denture base / Apex Flex — intaglio toward platform; anterior ~60°. Auto-orient is acceptable for most cases.
12
Tank II: Denture teeth / Apex Teeth — occlusal toward and parallel to platform.
13
Confirm supports: intaglio and interproximals clear of supports; adequate supports for success. Edit manually only if needed.
14
Send to the Pro 2 queue. Flex base must not be aimed at any printer other than Pro 2.
Fallback if Duo unavailable: sequential Flex then Teeth jobs at 100 µm — document in SEAZONA.

D. Print

15
Confirm Apex Flex and Apex Teeth are not expired.
16
Shake Apex Flex for 1 minute. Pour to at least the minimum fill line. Record lot in SEAZONA.
17
Shake Apex Teeth for 1 minute. Pour to at least the minimum fill line. Record lot. Stir/filter tank resin if already poured.
18
Install and lock Duo Kit adapter, platform, and tanks. Protect tank bottoms when off the printer.
19
Assign the correct material to each tank/side on the touchscreen.
20
Confirm the build platform is clean and free of debris or cured resin.
21
Optional: preheat resin. After preheat, start within 15 minutes or the printer will re-preheat.
22
Start print from the queue (typically ~32–40 minutes at 100 µm — trust printer status).
23
When complete, wait for the safety lid lock to disengage. Remove the Duo Kit platform carefully.
24
QC-B: Fully formed; no delamination; inspect before wash. Fail → remake.

E. Wash &amp; dry

25
Install the Duo Kit wash adapter in ProWash S. Seat the Duo Kit platform securely.
26
Run Standard Multi-Cycle Wash with IPA ≥ 91% (3 min wash / 3 min rinse / 3 min dry per Apex Flex IFU).
27
Keep teeth on the build platform during wash. Base may remain on the platform or wash in the basket.
28
Fully dry base and teeth with compressed air — especially tooth sockets, grooves, and internal surfaces. Residual IPA interferes with bonding and final cure.
QC-C: Clean and fully dry. Do not assemble or cure wet parts.
29
Remove parts for assembly. Leave supports attached to the denture base through assemble/tack and until after NanoCure.

F. Assemble &amp; post-cure

30
Roughen sockets or tooth bases if needed for bond.
31
Load Apex Flex into a syringe; apply into each tooth socket.
32
Seat teeth with firm, even pressure; tack-cure until teeth no longer move. Do not remove base supports before final cure.
33
Place the assembled RPD in NanoCure fully dry (supports still on). Select Apex Flex Denture and follow on-screen prompts.
If base and teeth are from different resin lines, select Conventional Denture instead.
Do not use classic ProCure for Flex bases.
34
After full cure, remove supports. Be careful around clasps. Prefer support snips; cut close to contact points. Remove the posterior support bar with snips.
QC-D: Cured, bonded, clasps intact. Fail → remake or salvage per remake criteria.

G. Finish, clean, bag (end of procedure)

35
Smooth remaining support contact points with a carbide bur at low RPM.
36
Polish using the SprintRay Master Workflow sequence: Resilience (B27/B29) → Tripoli (Robinson #11) → Ivoclar Universal paste (cotton buff) → mineral oil → fresh Robinson #11 cleanup. Low speed; keep wheels wet.
37
Clean with soap, toothbrush, and warm water (and/or steamer); dry. This is lab cleaning — not chemical disinfection.
38
Label and bag the prosthetic for outbound delivery from the in-house lab.
39
Complete the SEAZONA case log; mark status bagged / ready for outbound.
QC-E: Finish OK; cleaned; bagged; labeled; SEAZONA complete. Do not release if QC-E fails.
End of SOP-LAB-030. Office receipt, chairside disinfection before try-in, and patient delivery are handled under clinical protocols — not this manufacturing procedure.

9. Case Log (SEAZONA)

FieldEntry
Date
Patient / case ID
Medit Lab reference
Arch☐ Max   ☐ Mand   ☐ Both
Scan QC☐ Pass   ☐ Fail → returned
Designed by (Heygears)
Design approved by (Lab Technician)
Apex Flex shade + lot
Apex Teeth shade + lot
Printer / Duo usedPro 2 / Duo ☐ Yes
ProWash S cycle OK☐ Yes
NanoCure cycleApex Flex Denture ☐
Lab Technician initials
QC-SCAN + A–E pass☐ Yes
Remake?☐ No   ☐ Yes — reason: ________
Bagged for outbound☐ Yes — time ________

No SEAZONA bagged status = not released from the lab.

10. Quality Checkpoints

IDWhenPassFail action
QC-SCANAfter Medit Lab intakeScans usableReturn to office
QC-APre-printApproved design + minimaReturn to Heygears
QC-BPost-printComplete geometryRemake print
QC-CPost-washClean + dryRe-wash/dry
QC-DPost-cureBonded + curedRemake / salvage
QC-EBefore outboundFinish + clean + bag + SEAZONAHold release

11. Remake Criteria

FindingAction
Failed scan QCDo not manufacture — request new scans
Incomplete print / missing clasp / delaminationRemake print
Flex base printed on non–Pro 2Remake on Pro 2
Wet / under-cured / wrong cure cycleHold outbound; correct or remake
Teeth debond during finishRe-bond and re-cure if appropriate; otherwise remake
Torn or thin clasp after supportsRemake
Light support nubs onlySalvage — finish and re-check QC-E
UnsureStop and escalate to Doctor / lab lead

12. Competency

A Lab Technician may run this procedure unsupervised after:

RequirementSign-off
Read SOP-LAB-030 and Job Aid SOP-LAB-030-JATech __________   Date ______
Demonstrates Medit Lab intake, scan QC, Heygears portal submit/export/approve, and process controlsTrainer __________   Date ______
Completes 3 observed cases to bagged outbound with SEAZONA log1. ______   2. ______   3. ______
Trainer attests readyTrainer __________   Date ______
Lab lead authorizes unsupervised production__________   Date ______

13. Troubleshooting &amp; Escalation

ProblemAction
Bad scans in Medit LabQC-SCAN fail → return to office
Print failureRemake; filter resin; verify Duo / material assignment
Soft after cureConfirm dry part and Apex Flex Denture cycle; remake if needed
Teeth debondRe-bond or remake
Allergy or wrong indication notedStop — Doctor; do not manufacture

Escalate to Doctor / lab lead: unclear indication, allergy, chronic remakes, IFU deviation requests.
Escalate to office (Medit Lab / SEAZONA): failed scan QC; missing prescription.
SprintRay Support: 1-800-914-8004 — missing NanoCure Apex Flex Denture profile; repeated Pro 2 failures.