Smile Avenue Family Dentistry
IN-HOUSE LABORATORY Cypress & Katy, TX
Document ID: SOP-LAB-011
Version 0.1 DRAFT  |  Target effective 09/10/2026

ProWash S Maintenance

ProWash S care, IPA management, and station checks

Smile Avenue Family Dentistry — In-House Laboratory
SprintRay ProWash S — IPA, tanks, adapters, leaks

Document ID SOP-LAB-011
Version 0.1 DRAFT FOR REVIEW
Effective (target) 09/10/2026
Owner Lab Technician (process) / Doctor (clinical release)
Stack ProWash S with Smile Avenue Pro 2 platforms
Status DRAFT FOR REVIEW — do not train from this until Doctor sign-off

Related: SOP-LAB-001 (IPA grade, PPE, waste); product SOPs for wash cycle selection


1. Purpose

Maintain ProWash S so wash/dry cycles remove uncured resin reliably, IPA is replaced when saturated (not by guessing), tanks and filters stay clean, adapters are used correctly for Pro 2, and leaks are distinguished from normal evaporation.

2. Scope

In scope: IPA monitoring/replacement; reservoir tank scrub; mesh filter; hydrometers; platform vs basket wash; Pro 2 adapter rules; leak checks; drain-port issues; storage when idle.

Out of scope: Resin-specific custom wash times (see each IFU / product SOP); IPA hazardous-waste contracting (VERIFY vendor).

3. Who

Role Responsibility
Lab Technician Monitor hydrometers/prompts; drain/refill; clean tanks/filter; leak watch
Trainer / Lead Lab Technician Confirm Chamber 1/2 discipline; competency on drain hose
Doctor Approve drain-port kit / parts spend; clinical impact of under-washed cases

4. Safety

5. Materials

Item Notes
IPA ≥91% SprintRay recommended; ≥500 mL per chamber minimum on refill
Drain hose Pre-installed / clipped on unit
Cleaning brush (included) Scrub reservoirs when changing IPA
Mesh wash filter Clean when screen prompts
Wash basket Basket wash / combination workflows
Hydrometers Pre-calibrated per chamber — do not discard
Build platform adaptor Supplied for Pro55 / Pro55 Snot required for Pro 2 (Pro 2 magnet built-in)
Magnet sticker For older Pro95 / Pro55 detection — N/A for Pro 2 if magnet built-in
IPA waste container Labeled; never drain to sink
Drain port replacement kit If continuous leak / stuck port (SprintRay store kit)

6. Steps by cadence

6.1 BEFORE EACH WASH (daily ops)

  1. Confirm reservoir lids closed when idle; unit level, ventilated, nothing stacked on top.
  2. Check touchscreen for IPA / filter prompts.
  3. Glance at hydrometers: submerged ≈ OK; floating ≈ replace IPA.
  4. Select Platform Wash or Basket Wash per model (and IFU). For Pro 2: seat platform so unit detects it (built-in magnet — no Pro55 adapter).
  5. Do not stack/crowd models in basket; close basket lid fully when using basket mode.
  6. After cycle: inspect parts for residual resin; confirm dry (run Dry Only if moisture remains) before NanoCure.
  7. VERIFY Smile Avenue default cycle name(s) posted on station card vs resin IFU custom cycles.

6.2 WHEN TO CHANGE IPA (event-based — official SprintRay triggers)

Replace IPA when any of the following occur (SprintRay ProWash S IPA Care article):

  1. Touchscreen prompts replacement.
  2. Hydrometer floats.
  3. Cleaning results become inconsistent (sticky/tacky parts after a normal cycle).

Do not invent a fixed “every X days” interval unless practice later validates one from usage data — SprintRay guidance is condition-based. Chamber 1 (primary) typically saturates sooner than Chamber 2 (fine wash).

6.3 IPA DRAIN / REFILL / TANK CARE (each IPA change)

  1. Don PPE. Stage labeled IPA waste container.
  2. Drain (repeat per chamber as needed):
    - Remove drain hose from holder.
    - Press metal inlet into drainage port until it clicks.
    - Place free end in waste container.
    - Release safety clamp; drain fully.
    - Slide outer ring of drain port away to disengage hose.
  3. Clean IPA tanks every time IPA is replaced:
    - Remove hydrometer.
    - Scrub with provided brush.
    - Rinse with fresh IPA.
    - Reinstall hydrometer; repeat for second chamber.
  4. Add fresh IPA: open reservoir lid; add ≥500 mL IPA ≥91% to each chamber; close lid. (99% allowed; sensors may need several cycles.)
  5. Hydrometers may float briefly after refill, then sink once stabilized.
  6. Log date/chamber/volume (VERIFY log location).

6.4 MESH FILTER

  1. Clean when the screen prompts (SprintRay).
  2. Unscrew filter; rinse under cold water to remove debris; reinstall securely.
  3. Replace filter if it cannot be adequately cleaned (VERIFY spare filter on hand).

6.5 ADAPTERS / PLATFORM DETECTION (Pro 2)

  1. Pro 2 platforms: magnet is built in — do not require the Pro55 adapter or magnet sticker for detection.
  2. If “not detecting build platform”: reseat platform; confirm correct orientation; clean detection area; verify you are not using a bare Pro55 workflow by mistake.
  3. Pro55 adapter (if ever used on a loaner platform): place per SprintRay Pro55 adapter article — opening orientation as documented — VERIFY before use; Smile Avenue locked stack is Pro 2.

6.6 LEAKS VS EVAPORATION

  1. Slow IPA level drop with lids opened often / warm room → often evaporation. Keep lids closed; control temperature.
  2. Inspect quick-connect drain ports for drips beneath ports; confirm ports fully tightened and fittings lock.
  3. Normal: small drip on hose connect/disconnect.
  4. Abnormal: continuous leak, hose won’t lock, port stuck open/closed → drain chamber, clean, replace drain port with SprintRay kit (do not remove pre-applied Teflon tape; do not overtighten). If unsure, call Support before forcing hardware.

6.7 IDLE / STORAGE

7. QC

Check Pass
Hydrometers Submerged before starting production washes
Refill volume ≥500 mL/chamber documented
Tank scrub Done at every IPA change
Parts after wash No uncured film; dry before cure
Floor / ports No continuous IPA puddle

8. Troubleshooting

Issue Action
Floating hydrometer Drain/replace IPA; scrub tanks
Sticky parts Replace IPA; confirm correct cycle; check filter
Not detecting platform Reseat Pro 2 platform; clean; confirm no wrong adapter assumption
Level sensors odd after 99% IPA Run several cycles for sensor recognition
Continuous leak at port Tighten/inspect → replace drain port kit; escalate if persists
IPA disappearing, no puddle Evaporation controls; still inspect ports

9. Escalate

Escalate to When
Doctor Under-washed appliances that may affect fit/cure/biocompatibility; parts spend
SprintRay Support — 1-800-914-8004 Persistent leaks after port service, pump/sensor errors, detection failures after reseat, software/profile issues

10. Sources

11. VERIFY list

  1. Station default wash profiles vs each resin IFU custom cycle.
  2. IPA waste vendor, container labels, and pickup cadence.
  3. Spare mesh filter and drain-port kit on hand.
  4. Whether Chamber 2→1 recycle is used at Smile Avenue (SprintRay notes Chamber 2 can be recycled into Chamber 1 when needed) — practice rule yes/no.
  5. Log sheet for IPA changes (date, chamber, hydrometer/prompt reason).

DRAFT FOR REVIEW — Smile Avenue Family Dentistry · Cypress & Katy, TX · 09/10/2026